Tables Of Specifications - Sartorius IS...-X Series Operating Instructions Manual

Weighing platforms for use in hazardous areas/locations
Table of Contents

Advertisement

Tables of Specifications

Tables of Specifications
Model
SX
Readability
.........
Weighing capacity
Max. overload capacity
Tare range (subtractive)
Preload (can be electronically
compensated without reducing the weighing capacity)
Max. preload at start of isoTEST
calibration/adjustment
(platform must be zeroed)
Repeatability
Linearity
Sensitivity drift within +10 to +30 °C <±/K
Response time (average)
External calibration weight
(of at least accuracy class...)
Load plate dimensions
Platform dimensions (W x D x H)
Net weight,
approx.
Dust and water protection rating for the housing acc. to EN 60529
AC power
nician)
Frequency
Allowable ambient operating temperature
Operating temperature range
Adaptation to ambient conditions
Display update (depends on filter level selected) 0.1 - 0.1 - 0.2 - 0.4 (normal output rate)
Power consumption
Selectable weight units:
Built-in RS-485 interface
Model series Marking (ATEX)
IS.. BBE-.X..
h
II 2 G EEx ib IIC T4
IS.. CCE-.X..
h
II 2 G EEx ib IIC T4
IS... EDE-.X..
h
II 2 G EEx ib IIC T4
h
II 1 D T135°C (dust)
IS64EDE-HX..
h
II 2 G EEx ib IIC T4
IS... IGG-HX
h
II 2 G EEx ib IIC T4
h
II 1 D T135°C
* = factory setting
4
Operating Instructions IS-X
IS06BBE-SX
IS2CCE-SX
g
0.001
0.01
g
620
2200
kg
3
12
g
– 620
– 2200
g
93
110
g
110
1300
<±g
0.001
0.01
<±g
0.002
0.02
2 • 10
2 • 10
-6
s
1.5
1.5
g
500 (E2)
2000 (F1)
mm
Ø 130
218 x 200 218 x 200 218 x 200 218 x 200
mm
240x294x86
240x294x86
kg
7
7.3
source Power supply model YPS02-X or AC adapter model YPS02-Z (100-240V)
(must be connected by the weighing equipment dealer or a Sartorius Intec service tech-
Hz
48 – 62
0 °C to +40 °C (273 °K to 313 °K; 32 °F to 104 °F)
+10 °C to +30 °C (or 0 °C to +40 °C; depends on configuration)
By selection of 1 of 4 optimized filter levels
Average: 25VA
Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels,
Singapore taels, Taiwanese taels, grains, pennyweights, milligrams, parts per
pound, Chinese taels, mommes, Austrian carats, tola, baht and mesghal
(standard equipment; xBPI protocol):8 bits, odd parity; transmission rate: 9600 -
38,400 baud; half-duplex; convertible to RS-232 (SBI protocol): 7 bits, Parity: even,
odd, mark, space; transmission rates: 150 — 19,200 baud; 1 or 2 stop bits, software/
hardware handshake; factory settings: 1200 baud, odd parity, 1 stop bit, hardware
handshake with 2 characters after CTS
Use in
IP rating
Zone
1, 2 (gas)
IP 54
1, 2 (gas)
IP 54
1, 2 (gas)
IP 65
CL l, ZONE 1 AEx ib IIC T4
20, 21, 22 IP 65
IS CL II,III, DIV 1, GR E,F,G, T4
1, 2 (Gas)
IP 44
1, 2 (gas)
IP 65 and IP67
CL l, Zone 1 AEx ib IIC T4
20, 21, 22
IP 65 and IP 67
(dust)
IS6CCE-HX
0.01
6200
12
– 6200
5200
0.01
0.02
2 • 10
-6
-6
1.5
5000 (E2)
240x294x86
8.4
IP44
Markings in acc. with FMRC
approval for USA and CSA
approval for Canada
IS CL I, DIV 1, GR A,B,C,D, T4
CL l, ZONE 1 AEx ib IIC T4
IS CL I, DIV 1, GR A,B,C,D, T4
CL l, ZONE 1 AEx ib IIC T4
IS CL I, DIV 1, GR A,B,C,D, T4
IS CL I, DIV 1, GR A,B,C,D, T4
CL l, ZONE 1 AEx ib IIC T4
IS CL I, DIV 1, GR A,B,C,D, T4
IS CL II,III, DIV 1, GR E,F,G, T4
IS6CCE-SX
IS12CCE-
0.1
0.1
6200
12,000
25
25
– 6200
– 12000
1240
1200
6400
10000
0.05
0.05
0.1
0.2
4 • 10
4 • 10
-6
-6
1
1
5000 (F2)
5000 (F1)
240x294x86
240x294x86
7.3
7.3
Use in
Class I, Division 1 (gas)
Class I, Zone 1
(gas)
Class I, Division 1 (gas)
Class I, Zone 1
(gas)
Class I, Division 1 (gas)
Class I, Zone 1
(gas)
Class II,III Division 1
(dust+ fibers)
Class I, Division 1 (gas)
Class I, Zone 1
(gas)
Class I, Division 1 (gas)
Class I, Zone 1
(gas)
Class II,III Division 1
(dust + fibers)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents