Floor stand ovens

Heratherm floor stand ovens must always be attached to the wall using two (2) retaining
brackets on the outer left and right side on the back of the unit.
Remove the screws.
Attach the end of the retaining bracket that is facing downward to the unit.
Align the device at roughly 90°, +/-20° to the retaining bracket.
Affix the retaining bracket to the wall.
Install the anti-tilt anchor to a solid part of the building, which is able for shoring
loads.
The installation has to be carried out by qualified personnel only.
The connection to the building must be carried out with appropriate screws and
dowels according to the consistence of the building part.