Keysight Technologies N9029AV01-V15 User Manual page 7

Signal analyzer, frequency extension module and accessories

Advertisement

Always use the information printed on the SAX label of your module.
When using your signal analyzer with the SAX, you must set the analyzer to external mixing, including
harmonic and Doubler Switched parameters.
For the N90x0A Models:
[Input/Output] > External Mixer > External Mixer > Ext Mix Setup > Edit Harmonic Table >
[Enter] (or select OK).
Table Type > Doubler Switching > Harmonic > [N] > Enter N value (see SAX label on module, or see VDI
document, for N value)
For the N90x0B Models:
[Input/Output] > Select Input > External Mixer > External Mixer Setup > Table Type > Doubler Switching
> Harmonic (under Harmonic Table) > Enter N value (see SAX label on module, or see VDI document, for
N value)
For improved spurious response:
Change the number of sweep points:
[Sweep/Control] > Points > [40001]
Multi-Touch Sweep > Sweep Config > Points > [40001]
Reduce resolution bandwidth:
[BW] > Res BW > [300 kHz]
Turn on Signal ID:
[Input/Output] > External Mixer > Signal ID > On > Signal ID Mode > Image Suppress
Multi-Touch Input > Output > Signal ID > On > Signal ID Mode > Image Suppress
There is data on the USB drive that can be used to create a corrections file to improve amplitude
accuracy.
Keysight N9030-90061
When using the LO/IF (L) Input (External Mixing), use LO (L): N= X, with
"X" being the multiplication factor value shown on your SAX label.
When using the LO/IF (H) Input, use LO (H): N= X, with "X" being the
multiplication factor value shown on your SAX label.
SAX Label
On the SAX label, the multiplication
factor values are circled. This is an
example only; the actual values will
vary.
7