11. Remove the grommet that seals the blower's wire harness. (Fig. 7)
12. Remove the heater wire harness connector from the bottom of the chassis by
sliding the connector outward while pulling the top retaining tab upward slightly.
(Fig. 8)
13. Remove the 1/4" blower wrap foam, the blower, and the silicone isolator from
the underside of the chassis.
14. Carefully spray, wipe or soak the air intake, fl ow path and blower chassis inside
and outside using a 2% glutaraldehyde solution.
15. Wipe the power cord and the outside of both halves of the PAP case with the
same solution.
16. Let all components fully dry.
17. Reassemble the unit using parts from the Disinfection Kit—DV51D-682.
NOTES:
Route the silicone tubing correctly, as shown, and ensure that there are no
kinks. (Fig. 9)
Ensure that the silicone isolator is fully seated in the chassis and the blower.
Ensure that the grommet is fully seated around the blower wire harness.
Route the wire harnesses properly, using the provided strain reliefs on the chassis.
(Fig. 11)
Secure the power supply board within the mount clips on the chassis before
replacing the unit covers. (Fig. 10)
7
8
9
10
11
LT-2027
9