Application
– After threading the tubing and closing the lever, insert the removal spike|needle of
the Sterisart NF system as described in the instructions for the Sterisart NF sterility
test system. Then attach the test liquid container (e.g., bag or bottle) by hanging
it on the support rod (bag) or by placing it in the holding ring on the support rod
(bottle). The holding ring accepts bottles of various sizes. The holder must be turned
so that the sterility test system tubing is not under tension after the spike|needle
has been inserted and so that the spike| needle cannot slip out of the container.
For testing ampoules, an ampoule breaker is also attached to the support rod.
Use the integrated knife to scratch the neck of the ampoule, then snap off the top.
The broken-off top of the ampoule falls into the collection container (which can
be emptied later).
– For the following pumping process, make sure that the Sterisart NF sample
containers are correctly positioned in the intended holders, and that the tubing has
been inserted without tension or kinking.
Important Note!
When the lever is opened, the Sterisart Universal pump is turned off
immediately by a safety function. In this event, the Sterisart Universal
pump can be restarted after adequately closing all components.

6.1 Operating the Sterisart Universal Pump 16420

All operating parameters of the Sterisart Universal pump 16420 are saved in
programs. Once you have switched on the pump, it will enter the first program
saved (default = "Manual"). Press the right-hand soft key Menü (German for "menu")
to access the list of programs (program menu).
The following programs are predefined:
1. Manual
see section 6.1.1
2. Manual Record
see section 6.1.2