A&D MC Series Instruction Manual page 55

Mass comparator
Hide thumbs Also See for MC Series:
Table of Contents

Advertisement

Filter depending on differences in the amount to deliver powdery and liquid
material
When the weight value is not stable even without load fluctuations
Increase the averaging range for the first moving average parameter ( f1-b ).
Increase the averaging time for the first moving average parameter ( f1-t ).
Strengthen the digital filter. (Increase the function setting "Condition ( Cond )" parameter.)
When response is slow during full and medium flow
Decrease the averaging range for the first moving average parameter ( f1-b ).
Weaken the digital filter. (Decrease the function setting "Condition ( Cond )" parameter.)
When response is slow during dribble flow
Decrease the averaging time for the first moving average parameter (f1-t).
Weaken the digital filter. (Decrease the function setting "Condition ( Cond )" parameter.
To increase stability without load fluctuations
Increase the averaging range for the second moving average parameter ( f2-b ).
Increase the averaging time for the second moving average parameter ( f2-t ).
53