Class
Item and Parameter
at-f
Auto feed
dout
info
Data output
GLP output
ar-d
Zero after output
bp5
Baud rate
btpr
Data bit, parity bit
Crlf
Terminator
5if
Serial
interface
type
Data format
t-Up
Timeout
erCd
AK, Error code
Ct5
CTS, RTS control
Unit
Unit
d5 fnc
ld in
Density
Liquid density input
function
C5 in
Internal mass value correction
id
ID number setting
0
Not used
1
Used
0
No output
1
AD-8121 format
2
General data format
0
Not used
1
Used
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
0
7 bits, even
1
7 bits, odd
2
8 bits, none
0
CR LF
1
CR
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
5
CSV format
0
No limit
1
1 second
0
No output
1
Output
0
Not used
1
Used
g
gram
0
Water temperature
1
Liquid density
Refer to "7. CALIBRATION".
Refer to "10. ID NUMBER AND GLP REPORT".
Factory setting
31
Description
Selects whether or not automatic
feed is performed.
Selects GLP output method.
For how to set time and date to
be added, refer to the GX series
instruction manual, "10-9 Clock
and Calendar Function".
Adjusts zero automatically after
data is output.
CR: ASCII code 0Dh
LF: ASCII code 0Ah
Refer to the GX series instruction
manual, "10-6 Description of
Item "Data Format".
Selects the wait time to receive a
command.
AK: ASCII code 06h
Controls CTS and RTS.
Refer to the GX series instruction
manual, "5. WEIGHING UNIT".
Available only when density mode is
selected. Refer to the GX series
instruction manual,
"15. DENSITY MEASUREMENT".
Displayed only when the internal
mass value correction switch is
set to 1.