5 Confirm that there is nothing on the pan and press the
PRINT key. The balance measures the zero point and
displays the measured value. Do not allow vibration or
drafts to affect the balance.
The balance displays the target weight value.
6 Place the displayed target weight on the pan and press
the PRINT key. The balance measures the target weight
and displays the measured value. Do not allow vibration
or drafts to affect the balance.
7 The balance displays
pan.
8 The balance displays
Test Report" using the RS-232C interface or stores the
calibration test data in memory. For details on the
calibration test report format, refer to "10-2 GLP Report".
9 The balance will automatically return to the weighing
mode.