A&D MC Series Instruction Manual page 10

Mass comparator
Hide thumbs Also See for MC Series:
Table of Contents

Advertisement

Glass breeze break installation procedure
1. Remove the weighing pan, breeze ring, pan support
receptors, and dust guard from your balance.
To remove the dust guard, remove the two screws that
secure it to the balance.
Weighing pan
Breeze ring
Dust guard securing
screw (2 pieces)
Pan support receptor
(4 pieces)
Dust guard
Main unit
3. Secure the breeze break to the main unit using two
screws provided with the breeze break.
Replace the pan support receptors and weighing pan.
Weighing pan
Pan support receptor
(4 pieces)
Breeze break
securing screw
(2 pieces)
Rear left
Shock absorber
Front right
Caution: During installation, do not apply excessive force to the balance.
2. Place the breeze break on the main unit
Breeze break
• When installing the pan support
receptors at the front right and
rear left, note the following:
Push in the pan support receptors so
that the shock absorber springs are
above the receptor head surface. If
the pan support receptors are hard to
push in, hold the head of the receptor
as shown below and push it in while
rotating the head to the right and left.
• Keep the removed dust guard,
dust guard securing screws (2
pieces) and breeze ring.
• To reinstall the dust guard after
removing the breeze break, use
the two screws to secure the dust
guard.
• The dust guard securing screws
and breeze break securing screws
are different in length.