DIP Switch Settings
The M200 contains DIP switches on the back of the monitor allowing
you to configure various features by manually toggling the switch
from "A" to "B." See the table below.
1
See "Recording
Custom Pad Alarms"
on page 14.
A
Disable the use
of the recorded
message
– A user can not
record or listen to a
recorded message,
and the message
will not be played.
B
Enable the use
of the recorded
message
– Play the recorded
message followed
by the selected
alarm tone.
M200 Fall Monitor – Setup & User Guide
2
See "Latching and
Non-Latching" on
page 16.
Non-latching
– Local Audio and
Visual alarms are
reset when the
patient/resident
returns to the
pad.
Latching
– The alarm con-
tinues even if the
patient/resident
returns to the
pad.
Monitor Setup
3
See"Suspend Mode
(2 Minutes)" on
page 24.
Disable Suspend
mode
Enable Suspend
mode
4
See "Interfacing
with a Nurse
Call System" on
page 30.
Normally Open
(NO) Nurse Call
system.
– The NO system
looks for a
switch to close
for an alarm.
Normally Closed
(NC) Nurse Call
system.
– The NC system
looks for a
switch to open
for an alarm.
15
Need help?
Do you have a question about the M200 and is the answer not in the manual?