2. Enter the optic test page and select empty
cuvette.
Piercing Needles
Update Program
Vender
Prev
Maintenance
Empty Cuvette
Prev
Maintenance
3. Wait for 15 min and press "Start "to perform
the test.
4. The screen will show the measurement value
after the test finished.
5. Compare the data with the value in inspection
report. The tolerance range is as the table
shown below.Compare the data with the value
in inspection report. The tolerance range is as
the table shown below.
Wavelength T
260 nm
280 nm
320 nm
6. Adjust the optical system as shown in the flow
chart according to the count value.
83
Alarm History
System Status
Optic Test
Back
Reference
Back
olerance
±750
±1250
±1250
Standard Cuvette Vaule
S16C001514196
Integration time : 8 ms
Wavelengh
260mm
280mm
320mm
S16C001614196
Integration time : 8 ms
Wavelengh
260mm
280mm
320mm
S16C001714196
Integration time : 7ms
Wavelengh
260mm
280mm
320mm
S16C001814196
Integration time : 7ms
Wavelengh
260mm
280mm
320mm
If the value is out of tolerance,
adjust the integration time as next
Value
10437
18107
17475
Value
10291
19481
18124
Value
10172
20006
17944
Value
9158
19829
18786