Replace The Ejector Rod Module - RBCBioscience MagCore HF16 Plus Training Manual

Table of Contents

Advertisement

Replace the ejector rod module.

Open the front door.
1. Unplug the wire connector of the LED.
2. Loosen the three screws on the top of piston module (2.5mm Hex wrench).
3. Screw off the eight screws of the acrylic nameplate at the front side (3mm Hex
wrench), and take off the LED module.
4. Screw off the four screws of the reinforced metal sheet (2.5mm Hex wrench).
5. Screw off the three screws of the axle core of the ejector rod (2.5mm Hex wrench).
6. Take out the ejector rods downward.
7. Replace new ejector rods and reinstall the ejector rod module orderly.
2
1
4
3
6
5
41