Replace the timing belt of the lift motor of the base
Open the front cover and take out the waste box, cartridge rack and T-rack.
1. Unplug the power connector of the lift motor.
2. Screw off the six screws of the liquid container above the heating module (2mm Hex
wrench).
3. Screw off the two screws of the reinforced metal sheet of the motor (3mm Hex
wrench).
4. Screw off the two screws of the bearings at the lateral grooves (4mm Hex wrench).
5. Screw off the four fixation screws of the bearing fixation bases and move the drive
module upward (3mm Hex wrench).
6. Replace a new timing belt.
7. Note that the belt should be a little tight.
8. Install the lift motor, timing belt and timing gear on the machine.