Replace the lift motor of the base

Open the front door and take out the waste box, cartridge rack and T-rack.
1. Unplug the power connector of the lift motor.
2. Screw off the six screws of the liquid container above the heating module (2mm Hex
wrench).
3. Screw off the two screws of the reinforced metal sheet of the motor (3mm Hex
wrench).
4. Screw off the two screws of the timing gear (2.5mm Hex wrench).
5. Screw off the four fixation screws of the motor (2.5mm Hex wrench).
6. Replace a new lift motor.
7. Install the new lift motor orderly on the machine.
8. Note that the belt should be a little tight after installation.