Fuse Replacement At The Battery:

(Fig.4.5 4.6).
If your e-bike provides no motor support and the display
is blank, check the fuse at the battery.
Proceed as follows:
• Open fuse cover, (A).
• Pull out and check the fuse, (B), replace it, if
necessary
• Insert the fuse again
• Close fuse cover
CAUTION!
Only use the same fuse model with the same electrical
values (15A/80V).

Cleaning the battery:

CAUTION!
When cleaning the device, never use cleaning benzine,
thinner, acetone or similar agents.
Instead, use only conventional household cleaning
agents and disinfectants (eg.isopropyl alcohol).
The plug of the charging cable on the battery charger [9]
and the charger socket [3] on the battery may only be
cleaned using a dry cloth.
The battery must never be sprayed with a steam cleaner
or similar device.

Battery Storage:

We recommend storing the unused battery in a dry place
at temperatures between 10°C and 30°C.
• Do not keep your battery in places where the
temperature exceeds 45°C or is less than -20°C.
• Avoid direct sunlight.
• Avoid storage in humid areas to prevent corrosion on
the plug contacts.
• When the battery is not in use, recharge it at least
every 12 weeks to prevent damage to the battery."