Troubleshooting

Fault
❑ No aspiration.
❑ Pressure control
device switches fre-
quently.
page 30 of 35
Possible cause
➨ Filter clogged?
➨ Leak at the bottle head? ✔ Check seals, coupling
➨ No vacuum supply con-
nected?
➨ No vacuum while vacu-
um supply connected?
➨ Leak in the system?
➨ Filter clogged?
➨ Vacuum controller con-
taminated inside or
defective?
Remedy
✔ Replace filter.
and blind cap. Screw in
or replace if necessary.
✔ Connect vacuum supply.
✔ Check vacuum supply.
Open lock valve in the
vacuum line if applica-
ble.
✔ Check tubing, seals,
coupling, and blind cap.
Screw in or replace if nec-
essary.
✔ Replace filter.
✔ Contact local distributor.