5. Check Battery Charger Operation

a. Connect the common lead of the digital voltmeter to
the second pin connector on the power supply board.
Connect the other lead of the digital voltmeter to the BC
test point. See Fig. 5-2.
b. Vary the line voltage from 90 to 250 Vac w ith the
instrument on or off.
c. Check fo r a reading of +0.6 to +0.8 o f a vo lt on the
digital voltmeter.
d. Check fo r a 0 volt reading on the digital voltmeter,
when the instrument is disconnected from the power line.
(Behind Pot)
TP
R644
0 V TP
Low Voltage
Fig. 5-1. Location of the Low Voltage, High Voltage adjustments
and test points.
Fig. 5-2. Location of the +75 volt, +48 volt and the battery charger
test points.
R662
High Voltage
- 1 0 0 0 V
TP
REV. APR 1974
Calibration—221 Service
+75 Volts 5th Pin Connector From Rear
+48 Volts 4th Pin Connector From Rear
2nd Pin Connector From Rear
OANC
i
Battery Charge (BC)
Test Point