Prelim Inary Procedure; Prelim Inary Control Settings - Tektronix 221 Instruction Manual

Hide thumbs Also See for 221:
Table of Contents

Advertisement

Calibration—221 Service
Description
7. Cable
Impedance,
RG-58/U, length, 42 inches,
connectors, BNC.
8. Adapter
Connectors, BNC female and
dual banana plug.
9. Adapter
Connectors, GR874 and BNC
female.
lO.Termination
Im p e d a n c e ,
accuracy,
BNC.
11. Attenuator
Ratio, 10X; connectors, BNC;
impedance, 50 ohms.
12. Adapter
Connectors,
BNC.
13. T-Connector
Connectors, BNC.
-14. Screwdriver
Three-inch
bit.
3/4-inch shaft.
15. Low-Capacitance
Screwdriver
5/64 inch bit.
P R E LIM IN A R Y PROCEDURE
This instrument should be calibrated at an ambient
temperature of +25°C (±5°C) for best overall accuracy.
1 Disassemble as instructed in the Maintenance Section.
2. Connect the instrument to 90 Vac, to 250 Vac, 48 Hz
to 62 Hz or 80 Vdc to 250 Vdc line source.
3. Set the instrument controls as given in the Prelimi­
nary Control Settings. A llow at least five minutes of
warmup before proceeding.
NOTE
Titles fo r external controls o f this instrument are
capitalized in the procedure (e.g., INTENSITY).
Internal adjustments are in itia l capitalized only (e.g..
Sweep Cal).
5-2
Table 5-1 (cont)
TEST EQUIPM ENT R EQ U IR E D
Minimum Specifications
50 ohms; type,
5 0 o h m s ;
2%,
connectors,
probe
tip
to
shaft, 3/32 inch
Usage
External trigger operation check.
H o rizo n ta l
gain
check
adjustment.
External trigger checks.
Vertical Am plifier compensation
checks and adjustment
Vertical Am plifier compensation
checks and adjustment.
Vertical
Am plifier
bandwidth
check.
Used throughout procedure for
signal interconnection.
E x te rn a l
trig g e r
operation
checks.
Used throughout procedure to
adjust variable resistors.
U sed
to
a d ju s t
v a ria b le
capacitors.
4. See the Adjustment Locations in the pullout pages.
P R E LIM IN A R Y CONTROL SETTINGS
Preset the instrument controls to the settings given
below, when starting the calibration procedures.
VO LTS/DIV
SEC/DIV
Vertical POS
Horizontal POS
TRIGGER
TRIG LEVEL/SLOPE
INPUT COUPLING
HORIZ-MAG
VAR SEC/DIV (Time/Div)
VAR VO LTS/DIV
INTENSITY
Examples of Applicable
Test Equipment
T e k tr o n ix
P a rt
and
012-0057-01.
T e k tr o n ix
P art
103-0090-00.
T e k tr o n ix
P a rt
017-0063-00.
T e k tr o n ix
P art
011-0049-01.
T e k tr o n ix
P a rt
011-0059-02.
T e k tr o n ix
P art
013-0084-01
T e k tr o n ix
P art
103-0030-00.
Xcelite R-3323.
T e k tr o n ix
P a rt
003-0675-00.
J.F D
A d justm e nt
5284.
0.1
1 m
Midrange
Midrange
Internal
AUTO PRESET
GND
IX
CAL
CAL
(Out of CCW Detent, ON)
As desired.
REV. B, AUG. 1976
Number
Number
Number
Number
Number
Number
Number
Number
Tool

Advertisement

Table of Contents
loading

Table of Contents