f. Set VOL TS-mA/DIV to .2 (black scale).
g. Set the Generator frequency to 10 kHz and
its
out-
put amplitude for a 6 division crt display.
h. Set the Generator frequency to 1 MHz.
Calibration-213 Service
i. CHECK-Vertical deflection is at least 4.2 divisions.
j. Disconnect the Generator.
6-9