Characteristics
2 V to 100 V/div
Max Safe Input Current
DC Current Rejection
1
Input Leakage Current
1
Trigger Sensitivity
Intl AC Norm
Intl AC Auto
Ext DC
Trigger Jitter
External Trigger
Input Resistance
Input Capacitance
Maximum Usable Input Voltage
1
Maximum Safe Input Voltage
1
Sweep Rate
Calibrated Range
Accuracy Over
Center 8 Div
TABLE 1-2 (continued)
Oscilloscope Electrical Specifications
Specifications-213 Service
Performance Requirement
VERTICAL DEFLECTION SYSTEM (cont)
800 V (de+ peak ac)
2 A rms on any range, 3 A peak
No provision for blocking de current for ac only reading
200 pA maximum at 30°C or less (0.4 div or less trace
shift from open to shorted probe)
TRIGGERING
0.5 div, 7 Hz to 1 MHz
Same as Intl AC Norm except the sweep free-runs and
provides a bright reference trace when the vertical input
signal is not sufficient to cause triggering
1 V, de to 1 MHz
0.5 µs or less at 1 MHz
1 M.Q
±
10%
Approximately 20 pF
±
8 volts peak at 1 MHz or less
200 V (de+ peak ac)
HORIZONTAL DEFLECTION SYSTEM
500 ms
to
2 µs/div
Within 5%
Linearity of Any 2 Div Portion Within Center 8 Div
Within 5% {magnified and unmagnified)
Variable Magnifier Range
Horizontal Position Range
1
1
Supplemental information only.
@
Continuously variable between calibrated settings. Extends
fastest sweep rate to at least 0.4 µs/div. Each sweep rate
magnified by at least 5: 1
Start of sweep can be positioned to the right of center; end
of sweep can be positioned to the left of center.