Check Rms Current Measurement Accuracy - Tektronix 213 DMM Instruction Manual

Oscilloscope with options
Table of Contents

Advertisement

TABLE 6-10
Alternate RMS Voltage Measurement Accuracy
VOL TS-mA/DIV
Amplitude
Setting
Calibrator
(blue scale)
Settings
Limits
0.1
0.2 V
This reading should
.
1
2V
be the same as the
Digital Voltmeter
10
20 V
for the same number
100
100 V
of digits
13. CHECK RMS CURRENT MEASUREMENT ACCURACY
@
NOTE
To verify the specified accuracy of therms
current ranges with the procedure below, re-
quires the use of the AC Calibration Standard
(see footnote 3 in Table 6-1 ).
If
all previous
calibration steps have been accomplished, each
circuit function involved in rms current
measurement has been checked and adjusted;
therefore, further measurement accuracy
checks may not be necessary for the instru-
ments intended use.
a. Set mA pushbutton in.
b.
Connect the AC Calibration Standard (Current con
-
figuration) to the mA-D. and common jacks using two
banana-plug to BNC adapters and a cable.
Cal ibration-213 Service
c. Set VOL TS-mA/D IV and the Standard as shown in
Table6-11.
CHECK-Reading is within limits shown in Table 6-11.
e. Disconnect Standard.
TABLE6-11
RMS Current Measurement Accuracy
VOLTS-mAIDIV
AC Calibration
Setting
Standard
(blue scale)
Setting
Reading
Limits
(set at 40 kHz)
0.1
0.200 mA
0.200
.1950 to .2050
1
2.00 mA
2.000
1.925 to 2.075
10
20.00 mA
20.00
19.25 to 20. 75
100
200.0 mA
200.0
192.5 to 207.5
1000
1000 mA
1000
0925 to 1075
END OF PROCEDURE