Basics
Mode
Favorites
The Favorites mode displays the predefined
mode
tests. These can be selected via the favourites
bar.
Single mode In Single mode, a predefined test is started by
closing the module flap. Several tests can be
started and performed in parallel.
Test
The Test scheduler mode enables the user to
scheduler
configure the time and weekday when a
predefined test should be performed. To
configure the Test scheduler mode, select
Manual tests can only be started when the user
is logged in.
Log-in mode Only logged-in users (level 2-5) can perform
tests in Log-in mode. The configuration enables
you to define whether a user should be
automatically logged off after the completion of
a test. To configure the Log-in mode, select
To select the test mode:
1. Select
.
2. Select the desired test mode.
3. If necessary, select the
(only possible with the Test scheduler and Log-in mode
test modes).
4. If necessary, edit tests (see chapter 4.8.2 on page 12).
5. Save the settings with OK.
4.8.2
Creating or editing a test
NOTICE
i
i
A maximum of 3 favourites may be created.
The first 3 characters of the test name are displayed in
the list of favourites.
To create or edit a test:
1. Select icon
for the test to be edited.
The editing window will be displayed.
Enter test name
Select gas test
Further options
Further options
2. Enter desired test name.
12
Description
of the test mode to be edited
3. Select desired gas test.
The following gas tests are available:
-
Quick bump test
Extended bump test Test for reaching cylinder
Calibration
4. If required, select Options.
.
The following options are available:
Response time test
(only for calibration)
Bat. check
Alarm test
.
Zero check
NOTICE
i
i
When the alarm functions are being tested, the
ambient noise must not be too loud as otherwise the
station will not be able to check the horn function.
5. If required, select Further options:
Sync. clock
Download DL
Reset DL
Reset TWA
Turn off
Autom. repair
Fresh air
Generate
Print
1
Average shift values (time-weighted average) are generally
limited to eight hours exposure per day per workplace for
5 days a week during a work lifetime.
6. Save settings with OK.
No gas test, only the activated
options are run.
Test for exceeding A1
concentration.
concentration within a tolerance
range.
Zero point and span calibration
The response times of the sensors
are tested.
The battery voltage is measured
and displayed.
All the alarm functions of the gas
monitor are checked (noise, light,
vibration).
The zero point is checked.
The time on the gas monitor is
synchronised with the station.
If this option is selected, Reset DL
and Reset TWA are automatically
also activated to ensure that the DL
(data logger) is unique.
Download data memory and events
memory data from the gas monitor
to the station.
The data in the data and events
memories are cleared.
1
The TWA
time is reset.
The gas monitor is switched off
automatically after the test.
Perform automatic repair (e. g.
calibrate after a faulty bump test).
The system is purged with fresh air
after every test.
Create certificate (stored in the
station as a PDF file).
Print out the certificate on a
connected USB printer.