Maintenance; Maintenance Intervals; Performing A Firmware Update - Dräger X-Dock 5300 Technical Manual

Hide thumbs Also See for X-Dock 5300:
Table of Contents

Advertisement

9

Maintenance

9.1

Maintenance intervals

NOTICE
i
i
The maintenance intervals must be established in
each individual case and shortened if necessary,
depending on safety considerations, process
conditions, and the technical requirements of the
equipment. Dräger recommend a Dräger service
contract for all maintenance activities and that all
repairs are carried out by Dräger.
9.1.1
Before every start-up
The following work must be carried out before every start-up of
the equipment:
Check the hoses for dirt, brittleness and damage and
replace if necessary.
Check the hoses are secure, to prevent escapes of gas.
Check that all cable connections are secure.
Visual inspection of the modules and sensor seals. If very
dirty or if there are visible defects, the sensor seal must be
replaced.
9.1.2
Annually
Inspection of the entire X-dock station by competent
personnel.
9.2

Performing a firmware update

CAUTION
!
The station power supply must not be disconnected
during the installation process. The station may be
damaged if this is not observed.
NOTICE
i
i
The station does not support any USB data storage
device with an NTFS file system.
1. Download the firmware update from the internet:
a. Go to www.draeger.com.
b. Go to the X-dock product page and unzip the firmware
update to the root directory of an empty USB data
storage device.
CAUTION
!
There must not be any older firmware files on the USB
data storage device.
2. Connect the USB data storage device with firmware update
to the USB port on the station.
The USB icon will appear in the status bar.
Dräger X-dock 5300/6300/6600
3. Select
>
configuration > Update.
A list of all firmware updates
available on the USB storage
device will be displayed.
4. Select
the
desired
update from the list. The selected
firmware update is marked in blue.
5. Start the firmware update with OK.
The progress of the installation is
displayed:
6. Following successful transfer to the station, the start is re-
started automatically and the firmware update is installed
immediately afterwards. During the installation process,
the status LEDs on the modules will be white.
7. After the installation is complete, the station changes to
operating mode. The station is ready for operation.