Basics
When using a test gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with
Delete the current test gas component with
NOTICE
i
i
Deleting all test gas components deletes all
information of the test gas inlet.
2. Select test gas.
3. Enter the test gas concentration.
4. Select test gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the test gas cylinder
Batch number of test gas cylinder
Expiry date of the test gas cylinder
7. If required, enter Further options.
4.7.1
Further options
NOTICE
i
i
It is necessary to specify the volume, volume type,
pressure and unit to use the gas manager.
Configured test gas
components
Enter volume
Enter pressure
Status
Test gas pressure
1. Enter volume of test gas cylinder.
2. Set type of volume:
Comp. = cylinder volume
Relax. = gas volume (cylinder volume x pressure)
3. If required, refresh test gas pressure.
4. If necessary, activate Hose > 2 m if the length of the test
gas hose exceeds 2 m.
NOTICE
i
i
Dräger recommends not exceeding a hose length of
10 m for the gas feed hoses.
5. If required, configure other test gas inlets in the same way.
10
Status of test gas pressure:
Display
.
.
Gas cylinder level indicator:
Display
4.7.2
1. Select
Select test gas for X-am 2000
◄ Volume type
Fresh air inlet setting
Enter unit for
pressure
Expert settings
State of test
gas pressure/
Select test gas for PAC 7000 OV
Gas cylinder level
Gas cylinder level indicator
indicator
The OV sensor of the PAC 7000 OV allows to choose between
two different test gases that are also used for calibration and
testing. These test gases are carbon monoxide (CO) and
ethylene oxide (EO).
For the X-am 2000, it is possible to choose between three
different test gases that are also used for calibration and
testing. The three options are methane (CH
and pentane (C
sensitivity levels depending on the gas selected. More
information on this can be found in the relevant sensor data
sheets.
Gas pressure Meaning
<0.4 bar
Test gas pressure too low.
0,4 - 0,6 bar
Test gas pressure correct.
>0.6 bar
Test gas pressure too high.
Gas cylinder level
61 - 100 %
30 - 60 %
< 30 %
Gas cylinder level indicator deactivated
Settings
> Station gas configuration > Settings.
H
). The sensor is calibrated with different
5
12
Dräger X-dock 5300/6300/6600
), propane (C
H
)
4
3
8