Agilent Technologies E1339A User's And Scpi Programming Manual page 105

72-channel open collector digital output/relay driver module
Table of Contents

Advertisement

Code
Error Message
1500
External trigger source
already allocated
1510
Trigger source
non-existent
2000
Invalid card number
2001
Invalid channel number
2006
Command not supported
on this card
2008
Scan list not initialized
2009
Too many channels in
channel list
2010
Scan mode not allowed on
this card
2011
Empty channel list
2012
Invalid channel range
2600
Function not supported on
this card
2601
Channel list required
104 Relay Driver Error Messages
Potential Cause(s)
Assigning an external trigger source to a switchbox when the trigger
source has already been assigned to another instrument.
Selected trigger source is not available on this platform (e.g., some
triggers are not available on the Agilent E1300/E1301 VXI B-size
mainframes).
Addressing a module (card) in a swtchbox that is not part of the
switchbox.
Attempting to address a channel of a module in a switchbox that
is not supported by the module (e.g., channel 99 of a Relay Driver
module).
Sending a command to a module (card) in a switchbox than is not
supported by this card.
Executing an INIT before sending the SCAN command.
Attempting to address more channels than available in the switchbox
channel_list.
The selected scanning mode is not allowed with this module or you
have misspelled the mode parameter (see SCAN:MODE command).
No valid channels are specified in the channel_list.
Invalid channel(s) specified in SCAN <channel_list> command.
Attempting to begin scanning when no valid channel list is defined.
Sending a command to a module (card) in a switchbox that is not
supported by this card.
Sending a command requiring a channel_list without the
channel_list.