System Installation
58 • IL715 Tek-CARE
NC300II Manual
®
of approximately 26°C and relative humidity not exceeding 80%. If a PK250-series
battery backup is used, locate it in the same area. A PK305/A Supplemental Power
Supply is required with the NC351(A)/2 (as indicated above). Locate the PK305/A
near a dedicated 115 VAC outlet connected to the facility's Emergency Power Circuit
(i.e., connected to a backup power source), suitable for computer-grade equipment,
and within 2' of the NC351(A)/2. Interconnect the NC351(A)/2 and PK305/A using
the provided interconnection cable (CT305).
The PK305/A must always be powered down when the central equipment is
powered down. Do not stack equipment on top of the NC351(A), NC351(A)/2
or PK305/A. Do not locate these devices near other electrical systems, such
as fire alarm panels, electrical motors, air conditioning equipment, etc.
PK304, PK306 Master Station Power Supply: The PK304 provides +12 VDC power to
the NC304 or NC304LCD Master Station. The PK306 provides +16 VAC power to the
NC305LCD Master Station. The power supply connects directly to the master station,
and requires a dedicated 115 VAC wall outlet connected to the facility's Emergency
Power Circuit (i.e., connected to a backup power source), suitable for computer-grade
equipment (see
NC304, NC304LCD, NC305LCD and NC306 Master
NC304, NC304LCD, NC305LCD and NC306 Master Station: The NC304, NC304LCD,
NC305LCD or NC306 Master Station is used by the facility staff to interface with, con-
figure and program the nurse call system via the function buttons and numeric keypad.
It provides call indication on individual selector LEDs (NC306), a 2-line LCD display
(NC304LCD), a built-in LCD monitor (NC305LCD), or works in conjunction with an
NC313 VGA monitor (NC304 and NC304LCD) to display any or all system activity.
The staff may answer calls via the loudspeaker (controlled by the talk button) or by
using the handset for calls that require more privacy.
The NC304, NC304LCD, NC305LCD or NC306 is typically located on a desk or
countertop, within easy reach of the facility staff. The NC304, NC304LCD and
NC305LCD master stations must be located within 6' of a dedicated 115 VAC outlet
connected to the Emergency Branch of the facility's Essential Electrical System (i.e.,
connected to a backup power source), suitable for computer-grade equipment. (See
PK304, PK306 Master Station Power
6' of the included pre-wired wall plate, and it comes with a DB9 connector and cable
for interconnection. The pre-wired wall plate mounts on a single-gang ring or box,
and must be within 1000' of the NC351(A) or NC351(A)/2 Central Equipment.
The operating environment for the master station is approximately 26°C and relative
humidity not exceeding 80%. See wiring installation section for additional details. After
all connections to the master station have been made, please use the included ty-wrap to
secure all connected cables together for strain relief purposes. Battery backup is provided
using TekTone
PK250-series. If a PK250-series is used, locate it in the same area.
®
NC315 LCD/NC313 VGA Monitor: The NC315/NC313 is used in conjunction with the
NC304 or NC304LCD to display any or all system activity in standard English format.
The 15" monitor displays up to 22 events simultaneously. This display is also required
for system configuration and programming in conjunction with the master station. The
NC315/NC313 includes standard on/off, brightness and contrast controls, and a power
The NC315/NC313 is typically located on a desk or countertop, within easy viewing
indicator.
range of the facility staff. The monitor must be within 6' of the NC304 or NC304LCD
Master Station, and includes a DB15 connector and cable for interconnection. The
monitor includes a standard 3-prong connector and cable, and must be within 6'
of a standard 115 VAC computer-grade wall outlet. The operating environment for
the NC315/NC313 is approximately 26°C and relative humidity not exceeding 80%.
See

Wiring Installation

Supply.) The master station must also be within
section of this chapter for further details.
Copyright © TekTone Sound & Signal Mfg., Inc. All Rights Reserved.
Station).