Station Programming – By default, stations are assigned to the NC306 selector buttons
beginning with the lowest physical address on the lowest numbered port. The sta-
tions are assigned to the selector buttons sequentially, starting with the top button in
the left row (selector button 0), continuing down each row in sequence, and ending
with the bottom button in the right row (selector button 49). After all 50 selector
buttons have been assigned to stations, any additional stations are ignored—unless
an NC255 Master Station Expander is used to add 50 more selector buttons to the
NC306 (adding selector buttons 50 through 99).
The Program Station function allows the selector button assignments to be changed
from the defaults, so that a particular selector button can be associated with any
station on any port.
The station programming options are, in order: Assign Station to Selector Button,
Station ID, Bed Numbers, Station Labels Single/Dual Patient Stations, Audio Page On/
Off, Online/Offline, Privacy, Call Priority and Station Zones. (The last four functions
can also be programmed using the NC306 master's keypad.)
1. Type p, s, enter. The display will show "POINT (0–99)."
2. Type the number (0–99) of the selector button to be programmed, and press
enter. The LED on that selector button will start flashing, and the display will
show the physical address of the port and station assigned to that selector button.
3. Press enter to skip to next option, or type the physical address of the port and
station physical address to be assigned to this selector button, and press enter.
(Example: "07,15 enter" assigns the station on port 7 with dipswitch setting
15 to this selector button.)
4. Press enter to skip to next option, or type the 4-character alphanumeric room
ID to be assigned to the station associated with this selector button, and press
enter. (Station IDs are used by event printing, pagers, NC304-series masters
and NC305LCD masters.)
5. Press enter to skip to next option, or type the bed number(s) to be assigned
to the station, and press enter. (Example: "r, l, enter" changes the bed1 and
bed2 extensions of a dual station to R and L. The extensions are used by event
printing, pagers, NC304-series masters and NC305LCD masters.)
6. Press enter to skip to next option, or type a 1–16 character alphanumeric label
for bed1, and press enter. (Alphanumeric labels are displayed on NC305LCD
masters, and on NC304-series masters with VGA display monitors.)
7. Press enter to skip to next option, or type a 1–16 character alphanumeric label
for bed2, and press enter.
8. Press enter to skip to next option, or
• Type N and then press enter to turn Station Paging Off, or
• Type Y and then press enter to turn Station Paging On. (When station paging
is on, the station will receive audio pages.)
9. Press enter to skip to next option, or
• Type N and then press enter to take the Station Offline, or
• Type Y and then press enter to put the Station Online. (Take a station offline
when it must be switched off or physically removed from the system.)