Patient Station Operation - TekTone Tek-CARE NC300II Operation, Installation And Service Manual

Nurse call system
Table of Contents

Advertisement

System Operating Instructions

Patient Station Operation

Copyright © TekTone Sound & Signal Mfg., Inc. All Rights Reserved.
Method 1 – Disconnect the communication cable from the rear of the master station.
Taking the Master Station Offline:
Master fault indications and tones will occur on the other master stations. Cycling
power to the CE will clear the fault.
Method 2 – Using Master Station Button Presses:
• Press the talk (gray) button for 7–10 seconds, until the LED on the reset
(black) button is lit.
• Press the follow (orange) button. No fault signals will be generated and the
master station will retain its address information.
Method 1 – Disconnect the communication cable from the rear of the master, wait a
Bringing the Master Station Online:
few seconds, and then reconnect it. The master will be back online within one minute.
Method 2 – Switch off the power to the central equipment, and then power it up
again.
The Tek-CARE
NC300
Radio Paging Operation:
®
pocket paging applications via the serial ports. When properly configured, the radio
paging functions are transparent to the user and will happen automatically. The
details of setting this up are covered in
with the NC306 Master Station
The Tek-CARE
NC300
Event Printing:
®
a serial printer for event printing. When properly configured, the event printing
functions are transparent to the user and will happen automatically. The details of
setting this up are covered in the System Configuration and Programming chapter.
See the event printing sample in
This section refers to the following types of patient stations: IR301-series, IR302-series,
(Figure
10,
Figure
11)
IR311-series, IR312-series, IR318-series, IR319-series, IR320-series, and SF380-series.
Refer to
and
and indicators of these devices. A brief description of the operating controls for patient
Figure 10
stations follows.
Call-Placed Light: The call-placed light flashes when any call is placed from the patient
station.
Reset Switch/In-Use Light: The reset button cancels a call placed from the station. The
light illuminates whenever communications from the master station is engaged.
Call Cord Receptacle: There are three kinds of call cord receptacles. The 0.25" round re-
ceptacle, found on the IR301-series, IR302-series and SF380-series, uses SF301 and
SF302 type call cords. The 8P8C rectangular receptacle, found on the IR311-series and
IR312-series, uses SF301 pillow speakers, SF301DIG pillow speakers, SF311 call
cords or SF312 call cords. The round DIN receptacle, found on the IR319-series and
IR320-series, uses SF401EX strain relief cables, SF401 call cords, SF401P pillow
speakers, SF401DIG pillow speakers or SF41-series pillow speakers.
NOTE: Only call cords and pillow speakers that are tested with and UL
for use with TekTone
call cord receptacle on TekTone
properly isolated connections of hospital beds and other ancillary equipment.
II Nurse Call System provides support to interface with RF
in the System Configuration and Programming chapter.
Programming Pagers and Remote Stations
II Nurse Call System provides support to interface with
Figure
29.
for the locations, names and functions of the controls
Figure 11
nurse call equipment may be directly connected to the DIN
®
patient stations. Other methods are available for
®
IL715 Tek-CARE
approved
®
NC300II Manual • 19
®