ProMinent DULCOMETER D2C Operating Instructions Manual page 15

Adjustment and operation, measured variables ph/chlorine
Table of Contents

Advertisement

Operating Menu / Description
Type of limit transgression
Measured value 1 pH
Measured value 2 chlorine
Limit value
Measured value 1 pH
Measured value 2 chlorine
Limit relay 1 pH
Limit relay 2 chlorine
Limit relays 1,2
Switch-on delay* ∆t on
Switch-off delay* ∆t off
Hysteresis limits
measured vlaue 1
measured vlaue 2
Checkout time limits
Possible values
Initial value
Increment
upper
upper
lower
lower
upper
lower
pH 8
pH 0.01
pH 6.5
pH 0.01
0.1 ppm
0.01 ppm
0.6 ppm
0.01 ppm
LV 1
LV 1
LV 2
zone*
off
LV 1
LV 1
LV 2
zone*
off
active
active
closed
closed
active
open
0 s
1 s
0 s
1 s
pH 0.2
pH 0.01
0.04 ppm
0.01 ppm
off
1 s
Lower value
Upper value
pH -2
pH 16
pH -2
pH 16
0.00 ppm
upper limit
0.00 ppm
measuring range
0 s
9999 s
0 s
9999 s
pH 0.02
pH 16
0.02 ppm
2.20 ppm
1 s
9999 s
Remarks
Limit transgression
when exceeding or
dropping below value
*with regard to the setting
"zone", the difference between
the limits and the set
hysteresis should be ≥ 3x.
Is active in the direction
of cancellation of limit
violation.
Results in message and alarm
and shutting-off of the
corresponding metering.
Off: function off,
no message, no alarm.
15