Operating the Triad Freeze Dry System

Set-Up

Vacuum may be displayed in mBar, Pascal (Pa) or Torr and the temperature
may be displayed as either °F or °C. The refrigeration system and vacuum
pump operating time may be monitored and the operating vacuum level may
be set. To configure your Triad Freeze Dry System, turn the main Power
Switch ON. Press the DISPLAY button until the SET UP indicator is lit.
The display will show:
VACUUM UNITS:
MBar
 Press  or until the desired units are flashing.
 Press ENTER.
The display will show:
VACUUM SET POINT:
XXXX
Where YYY is the units selected above.
 Press  or until the desired vacuum operating level is displayed.
See "Setting the Operating Vacuum Level" for guidelines.
 Press ENTER.
The display will show:
TEMPERATURE UNITS:
^C
^F
 Press  or until the desired units are flashing.
 Press ENTER.
The display will show:
REFRIG TOTAL HOUR: XXXX
SERVICE HOUR: XXXX
Product Service 816-333-8811 or 1-800-522-7658
Chapter 4: Operating Instructions