4-12. Time Difference On Single Waveform; Rise Time - Tektronix 2215A Operator's Instruction Manual

Hide thumbs Also See for 2215A:
Table of Contents

Advertisement

INTENSI
FIED
/
ZONE
A TRACE
B
TRACE
r
-
ALT HORIZ DISPLAY
\
VERTICAL
REFERENCE LINE
C4207-23
Figure
4-12. Time difference on single waveform.
9. Adjust the B DELAY POSITION
control
clockwise to
move the
intensified
zone to the leading edge of the second
point of interest (on
the
A trace);
then
fine adjust until the
rising portion (on
the
B trace) is centered
at
the
same conve-
nient
vertical graticule used in preceding step 7.
10.
Record the B DELAY TIME POSITION control dial
setting.
11 . Calculate the time difference between repetitive
pulses using the following formula.
Time
(second
first
) (A
SEC/DIV)
Difference
=
dial
-
dial
switch
(Duration)
setting
setting
setting
EXAMPLE:
With the A SEC/DIV switch set to 0.02 ms
per divisions the first B DELAY TIME POSITION dial set-
ting is 1.20 and
the
second B DELAY TIME POSITION
dial setting is 9.53 (see
Figure
4-13).
Substituting the given values
in
the time difference
formula:
Time Difference
=
(9.53
-
1.20)
(0.2 ms/div)
=
1.666 ms
Operating
Procedures-2215A
Operators
DIAL READING
AT POINT
A
DIAL
READING
AT
POINT B
C4207-24
Figure 4-13. B DELAY TIME POSITION
control settings.
Rise Time
Rise-time measurements use the same methods as time
interval on single
waveform,
except that the measurements
are made between
the 1
0%
and
90%
points of the low to
high transition of of
the
selected waveform. Fall time is
mea-
sured between the
90%
and
10%
points of the high to low
transition of the waveform.
1 . Preset
instrument
controls and obtain a baseline
trace.
2. Apply an exact 5-division signal to either vertical-
channel input connector and set the VERTICAL MODE
switch to display the
channel
used.
3. Set the appropriate VOLTS/DIV switch and variable
control for an exact 5-division display.
4. Vertically position the
trace
so that the zero reference
of
the
waveform
touches
the
0%
graticule line and
the
top of
the waveform touches the
100%
graticule
line.
5. Set the A SEC/DIV switch for a single-waveform dis-
play. Ensure that the SEC/DIV Variable control is in the CAL
detent.
6. Select ALT HORIZONTAL MODE and set the
B
SEC/DIV
switch to spread the rise-time-measurement
portion of the display as much as possible.