Operators
Familiarization-
2215A
Operators
OPERATOR'S ADJUSTMENTS
INTRODUCTION
To verify the operation and accuracy of your instrument
before making measurements, perform the following
adjustment procedures. If adjustments are required beyond
the scope of the operators's adjustments, refer the
instrument to a qualified service technician.
Before proceeding
with
these instructions, refer to
"Preparation
for Use" (Section 2).
Verify that the POWER switch is OFF (button
out),
then
plug the power cord into the power-source
outlet.
If
indications
specified in these procedures cannot be
obtained, refer
the
instrument
to a qualified service
technician.
BASELINE
TRACE
First obtain a baseline
trace,
using the following
procedure.
1. Preset the instrument front-panel controls as
follows:
Display
A and B INTENSITY
Fully counterclockwise
FOCUS
Midrange
Vertical
(Both
Channels)
POSITION
Midrange
VERTICAL MODE
CH 1
BW LIMIT
Off
(button
out)
VOLTS/DIV
S0mV
VOLTS/DIV Variable
CAL detent
INVERT
Off (button out)
Input Coupling
AC
Horizontal
A/
8 SWP
SEP
Off midrange
POSITION
Midrange
HORIZONTAL MODE
A
A and B SEC/DIV
0.5 ms
SEC/DIV Variable
CAL detent
X10 Magnifier
Off (knob in)
B DELAY TIME
POSITION
Fully counterclockwise
B TRIGGER
SLOPE
LEVEL
A TRIGGER
VAR
HOLDOFF
Mode
SLOPE
LEVEL
A&BINT
A SOURCE
A
EXT
COUPLING
OUT
Fully clockwise
NORM
P-P AUTO
OUT
Midrange
VERT MODE
INT
AC
2. Press in the POWER switch button (ON) and allow the
instrument
to warm up (20 minutes is recommended for
maximum accuracy).
3. Adjust the A INTENSITY control for desired display
brightness.
4. Adjust the
Vertical
and
Horizontal
POSITION controls
as needed to center the trace on the screen.
TRACE ROTATION
Normally,
the resulting trace will be parallel to the center
horizontal graticule
line,
and the Trace Rotation adjustment
should
not
be required. If adjustment is needed, perform the
following procedure:
1
.
Preset instrument controls and obtain a baseline
trace.
2. Use the Channel 1 POSITION
control
to move the
baseline trace to the
center
horizontal graticule
line.
3. If the resulting trace is not parallel to the center hori-
zontal
graticule line, use small flat-bit
screwdriver
to adjust
the TRACE ROTATION
control
and align the trace with the
center horizontal graticule line.
3-3