Pro'sKit MT-1280 User Manual page 5

3 5/6 digital true-rms multimeter
Table of Contents

Advertisement

• Max. input volt drop: 200mV
• Max. input current: 20A (the test time should be in 10 seconds)
• 0.6A/250V Self-resettable fuse; 10A/250V fast-blown fuse
2-2-5.ACA
RANGE
60mA
600mA
20A
• Max. measuring volt drop: 200mV
• Max. input current: 20A (the test time should be in 10 seconds)
• 0.6A/250V Self-resettable fuse ; 10A/250V fast-blown fuse
• Frequency response: (40~1000)Hz
2-2-6. RESISTANCE (Ω)
RANGE
600Ω
6kΩ
60kΩ
600kΩ
60MΩ
600MΩ
• Open voltage: less than 0.7V
• Overload protection: 250V DC and AC peak value
NOTE: at 600Ω range, the test leads should be short-circuited, and measure the
down-lead resistance, then, subtract from the real measuring.
WARNING: DO NOT input any voltage at resistance range for safety concern!
2-2-7. CAPACITANCE (C)
RANGE
9.999nF
9.999nF~999.9uF
999.9uF~2000uF
• Overload protection: 36V DC or AC peak value
WARNING: DO NOT input any voltage at this range for safety concern!
ACCURACY
±(1.0%+5d)
±(2.0%+5d)
±(3.0%+10d)
True RMS
ACCURACY
±(0.8%+5d)
±(0.8%+3d)
±(1.0%+25d)
±(5.0%+10d)
ACCURACY
±(5%+35d)
±(2.5%+20d)
±(5%+35d)
4
RESOLUTION
10uA
100uA
10mA
RESOLUTION
0.1Ω
10Ω
100Ω
10kΩ
100kΩ
RESOLUTION
10pF
10nF
1uF

Hide quick links: