DeVilbiss 9352D Cleaning & Disinfection Manual page 2

Mask

Advertisement

Pasteurization Procedure
(Serenity ® and FlexSet Masks only)
1. Create a circulating water bath filled with sterile purified water (PURW) to 70°C (158°F).
2. Immerse the mask in the water bath and expose for 30 minutes.
3. Remove the mask from the water bath and air dry
Disinfection Procedure
1. Prepare a disinfectant compatible with Polycarbonate and Silicone according the manufacturer's
recommendations.
2. Immerse the mask completely, filling all areas and eliminating air pockets.
3. Following the disinfection exposure, thoroughly rinse the mask by immersing it com-
pletely in a large volume of purified water. Allow the mask to remain in contact with the
water for a minimum of one minute.
4. Remove the mask and repeat the rinse procedure a minimum of two additional times
using fresh, purified water for each rinse.
5. After rinsing, allow the mask to completely air dry before use.
NOTE: Follow all manufacturer's instructions and guidelines for product use.
D
e
V
l i
i b
s s
M
a
k s
Serenity Silicone (Standard 9352D and Shallow 9352S)
Serenity Gel (Standard 9352G and Shallow 9352GS)
FlexSet Silicone (Standard 9354D and Shallow 9354S)
FlexSet Gel (Standard 9354G and Shallow 9354GS)
Note: Velcro® patches may release from the FlexSet forehead pad with repeated disinfection.
(Replacement Forehead Pad: p/n 9353D-604)
®
VELCRO
is a registered trademark of Velcro Industries B.V.
© 2011 DeVilbiss Healthcare LLC 02.11 DeVilbiss, Serenity, and FlexSet are trademarks of DeVilbiss Healthcare. LT-1883 Rev E
DeVilbiss Mask Cycle Tolerances
DeVilbiss Healthcare LLC 100 DeVilbiss Drive Somerset, PA 15501 USA
800.338.1988
814.443.4881
.
l C
a e
n
n i
g
P
s a
e t
r u
26
10
26
10
26
10
26
10
www.DeVilbissHealthcare.com
z i
t a
o i
n

Advertisement

loading

This manual is also suitable for:

9352s9352g9352gs9354d9354s9354g ... Show all