To Test The Auxiliary Power - Agilent Technologies 1680 series Service Manual

Logic analyzer
Table of Contents

Advertisement

Chapter 5: Troubleshooting
General Troubleshooting

To test the auxiliary power

The +5 V auxiliary power is protected by a current overload protection device. If
the current on pins 1 and 39 exceed 0.33 amps, the circuit will open. When the
short is removed, the circuit will reset in approximately 1 minute. There should
be +5 V after the 1 minute reset time.
Equipment Required
Equipment
Critical Specifications
Digital Multimeter
0.1 mV resolution, better than
0.005% accuracy
• Use the multimeter to verify the +5 V on pins 1 and 39 of the probe cables.
112
Recommended
Model/Part
E2373A