GDD Instrumentation MPP3wifi Instruction Manual page 52

Magnetic susceptibililty, conductivity meter
Table of Contents

Advertisement

Step 10: Stamping and moving the probe from the second row to the third row
10. Once the probe is at the end of the second row, press MEM to stop the recording (or click on the
STOP icon of the MPP3wifi application).
11. Press STAMP to mark the 7m point (or click on the Stamp icon of the MPP3wifi application).
Stopping the recording before marking the last point of the row prevents recording too many
readings at the same place caused by many operations to do at the same time.
Step 11: Restarting the recording
12. See Step 1.
Steps 12, 13 and 14: Stamping
13. See Steps 2, 3 and 4.
Instrumentation GDD inc.
2017-08-21
Next point to
mark : 7m
Next point to
mark : 8m
51