Problem:
The message Please initialize the probe to continue appears during measuring process.
Solutions:
Move away the probe from any conductive or magnetic material (point it in the air) and try to
initialize the probe again by clicking on the INITIALIZE icon (see Section 10 – Toolbar).
Problem:
The message Server error (TIMEOUT) appears.
Solutions:
This message can appear if there is no activity after a while or if the MPP3wifi probe has been
turned OFF accidentally. Click on Retry to restart the MPP3wifi application.
Verify if your device is still connected to the MPP Wi-Fi
application and open it again.
If your device is no longer connected to the MPP Wi-Fi
no longer available, make sure that the MPP3wifi probe is ON and that the AA batteries are
sufficiently charged.
If none of these solutions suggested in the Section 15 – Troubleshooting can resolve your problem,
contact Instrumentation GDD –Technical Support (see Section 16 – Technical Support).
Instrumentation GDD inc.
®
network, then close the MPP3wifi
®
network or if the MPP Wi-Fi
2017-08-21
®
network is
43
Need help?
Do you have a question about the MPP3wifi and is the answer not in the manual?