System Configuration (Toolbar) :
Important : we highly recommend to warm-up the probe 50-60 minutes before taking measurement
to avoid any drift in the readings. If you can not warm-up the probe, set the Initialize time to 1
minute. If the probe has been warmed-up, set the Initialize time to 15 minutes.
All along the process, if the program ask to initialize, press on MEM on the probe to stop recording (or
click on the STOP icon of the MPP3wifi application). Then, move the probe away from any conductor
and initialize the probe by clicking on INITIALIZE. Then, put back the probe on the core samples and
press MEM to restart the recording (or click on Start (MEM) of the MPP3wifi application).
Step-by-Step process :
Step 1 : Starting
1. Move the probe away from any conductor and initialize the probe by clicking on INITIALIZE.
2. Place the end of the probe (sensor) on the core samples at the beginning of the row and press
MEM on the probe (or click on the Start (MEM) icon of the MPP3wifi application) to start the
recording and then, move the probe along the core samples. You need to press MEM and start
to move the probe right away to avoid having many recorded readings at the same place on the
core samples.
Instrumentation GDD inc.
(see Section 10 – Toolbar for more details)
2017-08-21
Next point to
mark : 1m
49