Opening And Closing The Housing; Rotating The Lcd Indicator - ABB TTF200 Operating Instructions Manual

Field-mount temperature transmitter. temperature transmitter for hart protocol. suitable for all standard requirements
Hide thumbs Also See for TTF200:
Table of Contents

Advertisement

16
TTF200 FIELD-MOUNT TEMPERATURE TRANSMITTER | OI/TTF200-EN REV. B
... 6 Installation

Opening and closing the housing

DANGER
Danger of explosion if the device is operated with the
transmitter housing or terminal box open!
While using the device in potentially explosive atmospheres
before opening the transmitter housing or the terminal box,
note the following points:
A valid fire permit must be present.
Make sure that no flammable or hazardous atmospheres
are present.
Figure 13: Cover lock (example)
To open the housing, release the cover lock by screwing in the
Allen screw
.
1
After closing the housing, lock the housing cover by unscrewing
the Allen screw
.
1
NOTICE
Potential adverse effect on the IP rating
Check the O-ring gasket for damage and replace it if
necessary before closing the housing cover.
Check that the O-ring gasket is properly seated when
closing the housing cover.

Rotating the LCD indicator

1 Front view
Figure 14: LCD indicator
The position of the LCD indicator can be adjusted to suit the
1
mounting position of the transmitter, to ensure that the display
is as clearly legible as possible. There are 4 positions at
increments of 90°-.
To adjust the position, proceed as follows:
1. Tighten the lock screw under the housing cover.
2. Release the housing cover by turning it counterclockwise.
3. Carefully pull the LCD indicator to release it from its
A11167
bracket.
4. Carefully insert the LCD indicator in the required position.
5. Screw the housing cover back on.
6. Loosen the lock screw until the housing cover is firmly in
place.
1
2
A11168
2 Rear view of LCD indicator / plug
positions

Advertisement

Table of Contents
loading