9.5.3 Controller information

RUN/ENTER

9.5.4 cal-test

The cal-test function is used to test the sealing of the system measuring
head - rubber sleeve - sample bottle and the operability of the system
®
OxiTop
Control.
In the BOD Standard mode with the GLP On setting, the controller indi-
cates when the next cal test is due, depending on the set calibration
interval (see section 9.4.4 GLP (
S
P
PECIAL AND
RESSURE P
To perform the test, you need the WTW test resource, OxiTop
order number 209 333.
Sample preparation: see operating manual of the OxiTop
resource.
Query stopped!
You can select between
- Continue query (see above)
and
- Stop (return to the CHECK
menu)
CHECK menu.
Use <><> to select the -
Controller Info menu item.
The display shows the following
information:
 Number of measuring heads
for which there is still memory
capacity
 Status of the supply batteries
 Status of the data backup
batteries (OK/LOBAT).
Use <><> to scroll to the
specification of the
 Software version
 Series number
(can differ from the example
shown here)
Return to the CHECK menu:
With <RUN/ENTER>.
OPERATING MODES
), page 61).
GLP/TOOLS
BOD S
, BOD
TANDARD
®
PM,
®
PM test
71