What to do if...
Display message
undef.
(Display in the curve pre-
sentation or sample sta-
tistics)
10 of 12
called up
(sample message, no. is
variable)
90
Cause
The selected measurement range
has been exceeded.
The sample filled was too warm.
The AutoTemp function is
switched off.
The sample filled was very cold
and has a low consumption
behavior (small BOD value).
The system is not sealed (bottle
internal pressure = atmospheric
pressure).
Two measuring heads have not
been found by the controller
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations)
The measuring heads have no
optical contact with the controller:
- Distance too great, angular posi-
tion imprecise
- IR window of the measuring
head averted from the controller
- Shading by other objects
The missing measuring heads are
defective.
®
OxiTop
Remedy
Select the correct measurement
range (see application reports on
the CD-ROM, Principles of measur-
ing technique).
Pre-temper the sample more pre-
cisely.
Switch on the AutoTemp function
(only effective with run durations of
more than 1 day).
Check the system for leaks by
means of a visual check:
Are the sealing surfaces of the bot-
tle and measuring head clean and
fault-free? Are there cracks in the
sleeve, bottle or measuring head?
Perform a cal-test (see section
9.5.4
-
, page 71).
CAL
TEST
Check the incubator temperature.
Modification of the distances or
angle.
Search at other storage locations.
Search for and check measuring
heads (see section 12.3 G
, page 93)