Prothrombin Time (PT)/INR Monitoring System Self-Test User's Manual
Coag-Sense
Meter Display
CLOT TIME
TOO SHORT
Page 40
Possible Cause
The clotting time was
very short and out of
testing range (<8
seconds).
An air bubble was
detected in the
sample.
For In Vitro Diagnostic Use
Solution
Repeat the entire
procedure (including
fingerstick on a different
finger) with a new strip.
Visually confirm that no
air bubbles are in the
sample before applying
to test strip.
Gently squeeze the
transfer tube bulb until
the blood exits being
sure not to introduce air
bubbles.
If the same message
repeats, contact
Technical Support.