10. Performing a PT Test

WARNING: Place the meter on a stationary, level
surface for testing. Do not move the meter or allow it to
vibrate during a test. Unreliable results may occur.
Action
1.
Make sure that the meter is
on by pressing the POWER
button on the top of the
meter.
Note ◙: If an error message
appears, refer to the
"Troubleshooting"
section.
2.
Open a PT test strip package, tearing at the notched end.
Remove the strip. Set the package aside.
Note ◙: Make sure that the expiration date has not passed.
3.
Holding the round end, gently push the strip completely into the
meter in one smooth motion. The strip fits snuggly when pushed
all the way toward the back wall of the strip holder.
Page 24
Meter Display
For In Vitro Diagnostic Use
READY
INSERT
STRIP
OFF
SET
MEM