Prothrombin Time (PT)/INR Monitoring System Self-Test User's Manual
Coag-Sense
Action
7.
Immediately after collecting
the patient sample, place
the tip of the sample
transfer tube at a 45º angle
into the sample well on the
test strip where you see the
flashing green light. Gently
touch the tip down onto the
sample well.
8.
Slowly squeeze the bulb
until the blood leaves the
tube being careful not to
introduce air bubbles into
the sample. Keep pressure
on bulb while you pull your
hand away to avoid back
suction of sample.
9.
Discard the sample transfer
tube in a biohazard waste
container.
10. When the sample is
detected, the meter display
looks like this.
Note ◙: If this screen is not displayed, either not enough blood
sample was applied or the sample had bubbles in it.
Remove the strip and retest with a new strip.
Page 26
Meter Display
For In Vitro Diagnostic Use
TESTING
PLEASE
WAIT