Commissioning
6.5 Commissioning via handheld programmer
Wizard complete
Setting range:
Description:
6.5.3.2
Requesting an Echo Profile
1. Enter
PROGRAM
mode
2. Accessing
features
56
BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wizard menu
when Quick Start is successfully completed.)
To transfer Quick Start values to the device and return to Program menu,
press DOWN
(Finish). Then press LEFT
mode.
Navigate to: Level Meter > Diagnostics (3.) > Echo profile (3.1.).
Press RIGHT
to request a profile.
①
TVT
②
Confidence
③
Distance from Low Calibration Point to vertical cross-hair
④
Algorithm: tF (trueFirst)
⑤
Distance from flange face to target
⑥
Echo
Use UP
or DOWN
feature becomes active.
Move a
Press RIGHT
cross-hair
Zoom into
Position the intersection of the cross-hairs at the center of that area,
an area
select Zoom, and press RIGHT
Update the
Select Measure and press RIGHT
profile
Return to
Select Exit then press RIGHT
the previous
menu
to scroll to an icon. When an icon is highlighted, that
to increase the value, LEFT
. Press LEFT
.
.
Operating Instructions, 06/2017, A5E34647946-AD
to return to Measurement
to decrease.
to Zoom out.
SITRANS LR560 (mA/HART)