3. Open the pull-down menu Actions – Scan Device to synchronize parameters (upload
parameters from the device to AMS).
4. Configure the device via the Quick Start wizard.
Quick Start Wizard via AMS Device Manager
Note
• A reset to Factory Defaults should be performed before running the Quick Start Wizard if
device has been used in a previous application. See Master Reset (Page 79).
• The layout of the dialog boxes shown may vary according to the resolution setting for
your computer monitor.
• At each step, you can accept the default values without modification and click on the next
step to proceed.
• After modifying parameters click on Apply inside the Quick Start window to write the new
values to the device.
• Click on OK only if you wish to update all parameters to the device and close AMS.
A Wizard provides an easy step-by-step Quick Start procedure that configures the device for a
simple application.
SITRANS LR560 (mA/HART)
Operating Instructions, 06/2017, A5E34647946-AD