Spo 2 Error Code Table - Cardiac Science 5300 Instruction Manual

Vital signs monitor
Table of Contents

Advertisement

Appendix
SpO
error code table
2
Sound
Error
Code
Priority
Type
E30
Medium
E31*
High
E32*
Medium
E33
Low
None
E34
Low
None
E35*
Medium
E36
Low
None
*For 5300, the alarm does not sound.
70-00582-01 A
A-10
Silence
Possible Causes
SpO
sensor not
2
Possible
connected.
Low signal level. Pulse
Possible
wave cannot be
recognized properly.
SpO
sensor has come off
2
Possible
patient.
Pulse signal detection in
-
progress
The current measurement
-
has been affected by
patient movement.
Outside of pulse count
Possible
measurement range
-
Pulse signal weak
Check Items
Sensor not connected. If
connected, the cable or connector
may be damaged. Replace with a
new cable. If replacing the cable
has no effect the problem may be
within the device. In this case,
cease use immediately and contact
Cardiac Science Technical Support.
Signal obtained from sensor is
weak. SpO
could not be
2
measured. There may be a problem
with fitting of the SpO
blood flow at the sensor site may
be unsatisfactory. Check the
condition of the patient and fitting
of the sensor, or replace the sensor,
and measure again.
Sensor is not in contact with
patient. SpO
could not be
2
measured.
Fit the sensor correctly to the
patient, and measure again.
Do not move the sensor mounting
location.
SpO
could not be measured due
2
to signal noise thought to be due
to body movement.
Ensure that the patient remains at
rest, and measure again.
Check the condition of the patient.
Also, do not move the sensor
mounting location.
Check the condition of the patient.
Also, check that the sensor is
correctly mounted.
Cardiac Science 5300/5350 Vital Signs Monitor