Appendix

Error code table

System error code table

Sound
Error
code Priority
Type
E01
High
E02
E81
High
E90
High
70-00582-01 A
A-2
Stop
Contents
ROM checksum error
Possible
RAM error
Possible
Battery is discharged.
Either no internal battery is
Possible
mounted or it cannot be
detected.
Check items
An error was detected in the main unit's
internal ROM. Switch the power for the
main unit OFF, then ON again.
If this does not solve the problem, there
is a possibility that the device is broken.
Immediately stop using monitor and
contact Cardiac Science Technical
Support.
An error was detected in the main unit's
internal RAM. Switch the power for the
main unit OFF, then ON again.
If this does not solve the problem,
immediately stop using monitor and
contact Cardiac Science Technical
Support.
An error was detected. Charge the
battery.
If this does not solve the problem,
immediately stop using monitor and
contact Cardiac Science Technical
Support.
Either no internal battery is mounted or
it cannot be detected.
Check the internal battery cable
connection.