Setting Method For Each Item - Cardiac Science 5300 Instruction Manual

Vital signs monitor
Table of Contents

Advertisement

Setup
Use the Forward or Back switch to change the contents of the setting
3.
Press the Menu/Enter switch to enter the setting contents.
4.
The setting content is entered and the display moves to the next setting item.
To exit Utility Mode and return to the basic screen, switch OFF the power for the
5.
main unit, then switch it ON again.

Setting method for each item

Measurement mode (Adult/Neonate) selection
1.
Adult: Set to this value when using reusable or disposable cuffs for adult and pediatric
patients.
Neonate: Set to this value when using the disposable cuffs for neonates.
For details, see
Adult initial inflation pressure value setting
2.
This is the inflation pressure value applied when the Cuff Interval and Smart
Inflation are set to OFF, the Cuff Start/Stop switch is pressed and the blood pressure
is measured. (5350 only)
For the 5300, this is the inflation value when Smart Inflation is set to OFF.
For details, see
70-00582-01 A
10-12
Forward switch
Back switch
Selecting the measurement mode
Initial inflation value
on page 4-16.
1
2
3
1
INI PRS is displayed.
2
The current initial inflation pressure value is displayed.
Display example: 180 mmHg
Available value: 140, 180, 220mmHg
3
The adult mode icon is lit up.
on page 4-4.
Adult/Pediatric
:
Neonate
:
Cardiac Science 5300/5350 Vital Signs Monitor