9. OPERATION OF CRYOSTAT

9.1 UVC DISINFECTION
UV-C radiation at 254 nm has an intense germicidal effect. Microorganisms, such as
viruses, bacteria, fungi and yeasts are effectively destroyed in the area accessible to UV
light without the addition of chemicals.
The cryostat TN50 can be equipped with a UVC light source that can either request
activation in defined periods (see section 7.3) or is manually activated.
The default setting for the UVC light duration and the automatic deactivation is set to
30 minutes. The timer can be set by an authorized service technician.
For safety reasons UVC light will be deactivated upon opening of the
i
lid of the cryochamber. The UVC disinfection will be aborted and
has to be activated again for surface disinfection.
The default setting for automatic deactivation is set to 30 minutes.
The timer can be set by an authorized service technician.
9.1.1 INSTALLATION INSTRUCTION FOR UV LAMP - CRYOSTAT
Please follow the respective steps and pay attention to the relevant instructions.
Never directly touch the UV lamp´s glass, since this reduces its durability.
Picture 1:
Operating Instructions – TN50
[only applicable to instruments equipped with this feature]